KOSMA 3-M
observing cookbook (from its Germany web site)
Jump Station: AOSs, bugs, catalog, chop, focus,
hexapod, pointing, skydip, tuning, 230,
345.,660,
0) Pumping and cooling down the receiver
(updated 20.10.01 by CK)
This procedure describes what to do in case the 230/345 GHz receiver
has warmed up above about 100 K. Some more details are found in the manual
`Bedienungsanleitung für die 4K-Maschine (2-Kanal-Empfänger)' which can be
found in a green folder at the Observatory.
START: (back to top)
| Turn to zero both mixer currents with L (=Mixer Bias) and at the E-Box.
Turn off both Gunns (C). Switch off both two junctions (E-Box).
| Switch off the Cold Head in case it is still running. Check that the
water cooler (Rückkühler and pump) and the ventilators in the Anbau are
still running. Switch off the Ionenzerstäuberpumpe. |
|
Evacuation of the vacuuum box of the
receiver: (back to
top)
| Put the Turbopumpstand on the caroussel near to the receiver.
| In the case it's cold in the dome, first warm up the pump with a fan
heater for about 15 minutes. Otherwise it won't start.
| Connect the tube of the pump with the connecting piece of the vacuum
valve of the receiver.
| Switch on the pre-pump (Switch 1) and evacuate the tube which takes
about a minute.
| Open the vacuum valve at the receiver.
| When the pressure is less than 1 mbar (check the analog display of the
E-Box) switch on the turbo pump (small buttons ON and START of the
red-box). If it is still too cold it won't start. Switch to level 2 at the
pre-pump.
| Open the JT-valve of the receiver by 8 turns by turning the knob
counter-clockwise.
| Wait until the pressure is less than 0.01 mbar. This may take several
hours or one night. |
| | | | | | |
Cooling down the receiver: (back to top)
| Before starting the compressors make sure that the drive of the expander
(cold head) is running.
| Switch on the compressor RW to the right of the JT compressor.
Occasionally it doesn't work and you have to push the RESET button first.
| Now switch on the JT-compressor. The pressure in the JT-Loop (check
pressure display at JT compressor) before switching it on should be 6.5
bar. After switching on the JT compressor, the pressure increases to
approximately 9 bar. |
| |
Final adjustments: (back to top)
| When the 20K-stage temperature drops below about 30 K, the JT-valve can
be shut by 7 turns.
| Switch on the power supply of the Ionenzerstäuberpumpe, but don't push
the START or ON button, since the pump starts automatically when the
pressure is less than 10-5 mbar. Don't be bothered by the
clicking noise this machine makes for quite a while when trying to start.
| After approximately 10 hours the 20K-stage and the 4K-stage have reached
temperatures below 20 K and the JT-valve can be closed step by step until
the final temeratures are reached, e.g. 13 K and 4.3 K, - but the valve
must not be closed completely. By closing the valve, the temperature of
the 4K-stage may increase for one or two minutes but then it will decrease
again.
| The temperature of the 4K stage can accurately be read from the red
digits - Voltages - at the receiver. For lower temperatures these become
higher. 4.3 K corresponds to 1.623. At the end of the green folder there
is a conversion table voltages to temperatures.
| If the pressure is between 10-6 and 10-7 mbar the
pump can be removed. First close the vacuum valve at the receiver! At the
end, the pressure will be about 4 10-7 mbar. |
| | | |
Notes: (back to top)
| In the ``Anbau'', check that the ventilators in the window and that of
the ``Rückkühler (on the left) are running. The small pump in the
background should also work. The big water cooler in the center is not
used anymore. The machine to the right is used for the
``Deckenklimaanlage'' in the computer room.
| In case you have closed the JT-valve too far, i.e. the JT-valve was shut
accidentally, you should close it completely with a few turns. Then open
it again a few turns until you here a ``blob''. After that, only a 1/4-1/2
turn is needed to bring it to the correct position.
| The Penning vacuum measurement instrument should not be switched on
until the pressure is below 10-3 mbar (check analog display of
the E-Box).
| If the Penning vacuum measurement instrument has failed, the Ionenzerstäuberpumpe
cannot start. Remedy: unplug the measurement instrument and plug it again. |
| | |
Warning: If this procedure is not clear,
consult the receiver group in Cologne or an expert observer before you start
an experiment with a dubious result.
1) Starting up the telescope (back
to top)
- It is imperative to restart the telescope control system (TCS)
first, before reopening the dome.
Follow the instructions attached to the Telescope Drive Rack if you need to
do this. In order to restart after a normal >STOPTEL<,
issue a >STARTTEL< first on
Jansky. Make sure the button on the Telescope Drive Rack in the AOS room is
set to maintenance closed (this should
be the normal setting, when all is shut down).
AFTER
the telescope has found its initial position - the prompt on Jansky will
return, the Telescope PC Display in the control room shows that the
telescope is tracking in 0,0 position - go upstairs and set the switch on
the Dome Drive Rack to Manual (again, this should be the normal setting,
when all is shut down).
Now go into the dome and open the dome with the controls
in the dome. Once the dome is open, set the switch on the Dome Drive Rack
back to Remote (if the fuse trips (the light goes off), go into the AOS room
to the Dome PC, switch to Local and reset the Fuse with F - switch back to
Remote). Also, set the remote controller on the Telescope Drive Rack in the
AOS room to observe before you go down.
Now, start up MOK on the Jansky.
If no >STOPTEL< had
been sent, you may directly start up >MOK<
and read your catalog with >CAT<.
Start the online displays on Jansky with >KOSMOS<
if it is aborted during observations and type >SHOWD<
on one of the VT510s to see the telescope control and parameter display
(source name, R.A & Dec. Az. & El., observing mode, sky and rest
frequencies, velocities, etc - use 1 or 2 to switch between receiver setups
or to refresh the display). Note the setting for the synthesizer (SYN) on
the display, which you need for tuning the receiver.
2) Preparing an observing run: (back
to top)
- Each observation needs a preselected catalog to
read the observing parameters from. Such a catalog can easily be created
from an existing catalog or be edited on one of the VT510s. Just type >CAT< and read in the corresponding
catalog. Exit the editor with >CTRL Z<
and save the new catalog (>CTRL Y<
aborts).
3) Tuning the receiver: (back
to top)
Each receiver will require a different procedure, but a few things are
common to all of them. First, make sure that the tuning
PC (it has a display in the control room and next to the telescope) is in
local mode. To achieve this, set the selection box in the dome to the terminal
you use (>local< for the dome
terminal, >remote< for the control
room terminal). Now type >F4< and
choose the receiver you want to tune and set it to >local< mode
(a tuning file for your frequency may be saved on the PC - ask around). Select
the receiver tune display with >F1<
or >F2< and follow the particular
tuning procedure:
230: (back to top)
| set the synthesizer frequency to the correct value (A).
| turn on the Gunn (C).
| set the Gunn Bias to approx. 9V (D).
| change the Gunn Tune (E) until the Gunn spike appears in the spectrum
analyzer (cf. RX 230
tuning table)
| lock the Gunn spike: use the PLL sweep (K) if necessary. Make sure you
lock the correct frequency, by changing the SYN frequency to another
harmonic frequency (SYN*HARM/new HARM), where it should remain locked.
| check the I(V) curve in the oscilloscope for the SIS photon step by
turning on and opening the mixer sweep on the bottom of the cryostat.
| now switch the BNC cable to see the conversion curve for the SIS
junction.
| move the Gunn Backshort (F) and the Multiplier Backshort Harm. (I) and
Fund. (H) to find and optimize the conversion curve. Try to set these in
such a way, that at first a double-peaked curve (a broad peak near the
center and a narrow peak on the right shoulder of it) appears in the
spectrum analyzer. Now optimize (F), (I) and (H) to totally suppress the
narrow peak on the right shoulder in order to get a symmetric curve with a
smooth curvature on top. Be aware that you may find a setting where you
can get a lot of LO power, but at the wrong frequency, i.e. the wrong peak
in the conversion curve is maximized. Make sure, the left peak is at its
maximum - even if that means operating at lower LO power. If moving (F)
unlocks the Gunn, move back the Gunn spike with (E). Note, that
increasingly higher values of (F), (I) and (H) will eventually collapse
the curve you try to optimize, but there are more than one maximum and you
should find the highest one for the left peak.
| minimize the sharp Josephson spikes to the right and left of the central
peak by changing the magnetic field strength (S).
| the peak amplitude of the curve can be increased by increasing the power
for the HEMT amplifier, but be aware that more than 2V may cause
interefrences, which will materialize as ripples in your 230 or 345 GHz
baselines.
| now turn down the sweep and change the Mixer Bias (L) to set the correct
working point to the maximum in the curve.
| check the Y-factor: it should be around 2. |
| | | | | | | | | | |
345: (back to top)
| set the synthesizer frequency to the correct value (A).
| turn on the Gunn (C).
| set the Gunn Bias to approx. 8V (D).
| now check the tuning
tables for the corrcect microscrew settings for the Gunn frequency and
the Gunn Backshort. The Gunn frequeny given is 1/3 of your actual LO
frequency. Set the microscrews <F> and <B> to the values given
in the tuning table. These are approximate values and you will have to move
<F> until the Gunn spike appears in the spectrum analyzer.
| lock the Gunn spike: use the PLL sweep (K) if necessary. Make sure you
lock the correct frequency, by changing the SYN frequency to another
harmonic frequency (SYN*HARM/new HARM), where it should remain locked.
| check the I(V) curve in the oscilloscope for the SIS photon step by
turning on and opening the mixer sweep on the bottom of the cryostat.
| now switch the BNC cable to see the conversion curve for the SIS junction.
| set the attenuator (see schematic) to maximum LO power.
| iterate between the Gunn and Tripler Backshorts to find and optimize the
conversion curve. The conversion curve should be symmetric with a smooth
curvature on top.
| minimize the sharp Josephson spikes to the right and left of the central
peak by changing the magnetic field strength (S).
| you can now move back the attenuator and try to optimize the shape of the
converison curve.
| don't touch the Harmonic Mixer Backshort.
| check the Y-factor: it should be around 2. |
| | | | | | | | | | | |
490: (back to top)
660: (back to top)
| to find the correct operation point for the doubler (tripler), first set
the corresponding I and V switches at the E-Box to the 660 doubler (tripler)
setting (note, that the switches can glitch, so count how often you click)
in order to read the current and voltage on the display. Now set the voltage
for the doubler (tripler) according to the table using the corresponding
labeled V potentiometer (e.g. 4V).
| if you now turn the micrometer screw of the doubler (tripler) at the LO,
you can see the current change on the LC display at the E-Box. Set doubler
and tripler to maximize the doubler(!) current (the tripler current is not
sensitive enough and is read only if you want to check the tripler's I(V)
curve). This should correspond to the maximum output power: make sure that
doubler and tripler are set to optimize the total output power when measured
directly with the voltmeter at the CONT output. |
|
Remeber to check the Y-factor and to set the system back to >remote<
before starting to observe.
- Make sure that the AOSs of your choice are connected to the correct
receiver. Switch the cables at the output of the IF box (right rack), if
necessary. Do NOT switch the cables at the IF Power Amplifier (left
rack) ! If you need to switch the connectors, make sure that the file >FAHR:[INC]PARAMETERS.MOK< is updated
accordingly. Check the water cooler and make sure it is set to 17.8-18 C and
that no air bubbles obstruct the flow of the water through the AOSs. Make
sure the AOS PC is set to 33 MHz.
In case you're using the VRS, make sure that
the resolution is set correctly. If you need to change the resolution,
follow the instructions attached to the VRS.
5) Starting the observations: (back
to top)
- First, you need to check the receiver temperature. Make sure, that there
is enough liquid nitrogen in the cold load and that the temperature for the
hot load is set correctly in MOK under >MENU
3<. This is also a good time to check if the weather
parameters have been updated automatically in >MENU
1<. Send a >LOAD<
and make sure that the AOSs bandpasses are not saturated during the hot load
measurement: if you need to lower the bandpass during the measurement,
repeat with >LOAD N< only. A
succesful load should now give you an acceptable receiver temperature.
Note:
| big spikes in the bandpass when using the 230 can occur at some
frequencies if the 345 Gunn is on
| even/odd noise can occur if the temperature of the AOS water cooler is
too high: try lowering it by 0.1-0.3 deg.
| |
| |
Now, you may want to point up the telescope on a planet followed by a
focus of the subreflector. For pointing on a
planet load the corresponding >PLANET.CAT<
catalog. Make sure that >AUPOI< is
running in MOK. The pointing is set up in such a way, that >GO< will map a cross around the planet,
giving you the pointing offsets in azimuth and elevation, which you may then
enter under >MENU 2< (enter >WAIT< after the beginning of each map, in
order to enter the pointing corrections before beginning the next map). To
do a focus, you need to load >PLANETP.CAT<,
which enters on the planet itself. Use e.g. >FOCUS
/Z 700 5/CENTER X< to take 5 measurements at X-700 mum, X-350
mum, X, X+350 mum, and X+700 mum. This should give you a reasonable fit for
the best z position (use >HELP FOCUS< for
more details).
Before you start observing your source, you need to do a calibration on the
sky. In principal, there are two ways to do this: the chopper wheel
calibration or the sky dip measurement.
Skydip:
| turn off the chopper wheel with >CHOP<.
| type >SKY< to load the
skydip catalog, choose the desired azimuth and do a >LOAD<.
| type >GO< to start the
measurement. |
| |
Chop:
| turn on the chopper wheel with >CHOP<.
| load an object catalog that uses Total Power mode.
| do a >LOAD< and send
the >GO<. |
| |
In case you continue to observe in (Dual) Beam Switch mode, make sure with
>CHOP< that the chopper is not
running before you enter the tau in >MENU
3< after the sky calibration. It's a good idea to add a
comment about the tau and weather conditions in the observation log with >REMARK "comment"<. Use >REMARK/URGENT< to write more important
comments into the observation log sheet.
Now you're ready to load your observing file with >CAT<,
send another >LOAD< and then start
the observations with >GO<. Make
sure the number of scans per calibration is set correctly in the load cycle
with >LOCY<. All spectra will be
written in CLASS format and you may want to open a new CLASS file for your
spectra with >NEXTFILE<. To stop
the observations after the current spectrum, type >FINISH<.
For an immediate stop, type >ABORT<,
but be aware that this may result in problems, if you abort in the middle of
an integration (do not abort a comb/dark/load measurement !).
In a seperate window on Jansky, you can run a >CLASS<
session after setting the data directory with >SET
DEF DATEN:< (you can set the size of the CLASS file indow with
>SET TERM/PAGE = 40<). The name of
the current catalog can be seen on the VT510 display. Just use the common
CLASS commands to read, reduce or print your spectra (>HA/PL PS_KOELN< will print a spectrum in Köln).
It's a good idea to frequently observe one
of the standard calibration sources listed in the binder you'll find at the
observatory. See if you get the correct integrated intensities and if so,
put a copy of the spectrum into the binder. This will help the next observer
and is a simple way to monitor the overall performance of the system.
|
Telescope in Elevation Limit:
If you stoptel the telescope without closing the dome, the telescope may hit
the low elevation limit. In order to drive it out by hand, deactivate the
computer control in the telescope rack, press the "rate" button
and then turn the "rate" knob above the buttons to "up".
If nothing happens, check if the green fuse behind the metal door on the
bottom of the telescope control rack (you can see it centered in the middle
without actually opening the door) is in! If it´s out, press it in (again,
you can reach it without opening the door), keep it pressed and now turn the
"rate" knob. After a few seconds the fuse will pull itself in and
stay in. Drive the telescope for about a minute, then set the control back
to the computer. Make sure the El is above 10 and Az above 0 for a starttel.
Otherwise the telescope will miss the 0,0 position and run into Az or El
limit again (from where you will have to drive it out again by hand).
|
Hexapod Move Error:
If this occurs, you can first try to reinitialize the hexapod motors by
resetting x,y,z to their current default values using >SUB< in
MOK and following the menu.
|
Wobbler Reset:
If you need to reset the Wobbler (see manual next to the wobbler control
rack) and follow the standard procedure "Reset Wobbler" -
"Start Wobbler" - "Start Wobbling" the wobbler may still
not move. Try setting the wobbler frequency (F) again last - now it should
start. Standard settings are 1800 for amplitude, -60 for position, positive
phase and frequency of 1 Hz. If you need to reset the Subref PC you need to
initialise the wobbler and then start MOK while the wobbler is Off and not
wobbling. Makes sure the values in the >SUBREF<
menu in MOK are ok as well. Here, the frequency setting should be
-1 Hz, i.e. extern!
|
Hexapod Leg stuck:
If one of the hexapod legs is in error (usually a value near -200 is being
reported), drive it manually "down" for some time (the readout
will not update and stay at -200, but the leg will move) and then wait
(maybe hours). To drive the leg from the Simatic rack, turn the computer
control off (push the button), flip the key and set the knob to the
corresponding leg. Try a few stoptels and starttels - eventually, the
correct readout will return and you can now drive the leg to its nominal
setting. When you´re near these settings use x,y,z and i,j,k in the other
menu (F1/F2) to find the true leg settings (they always vary by a few,
whereas x,y,z & i,j,k are absolute) - now continue observing.
|
Simatic Reset:
If you need to reset the Simatic rack, hold the little flip switch down at
the OR position. Now flip the Run/Stop switch twice and let go of the other
switch again - it will automatically flip back to the center position. You
have now reset the internal memory banks of the Simatic unit. If the Stop
light is now on, flip "only" the Run/Stop switch again. Now the
Simatic control is back (this procedure is also outlined in the Vertex
manual next to the Simatic unit). If you need to turn off the PC and it
doesn´t come back on, reset the red main power handle at the right bootm of
the Simatic rack.
|
OTF timing error:
Opening the CLASS windows on Jansky to display the OTF calibrated spectra
may cause a time error in the data transfer for short integration time
spectra. Close these windows with >F4<
to avoid time errors, if you have short integration times.
|
Jansky Prompt:
Don't type any messages following the MOK>
prompt without hitting <CR> (like remarks). This will block the
dispay for message outputs and thus halt your loop until the messages have
been written.
Suggestions, Improvements, Additions are welcome
by e-mail
[created 1999-10-12 by
atieftru@ph1.uni-koeln.de; updated 2000-11-11]
|
| | | | | |
(back to top)
|